Device recall: Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 / D150GC-40; Digital Radiography System RADspeed Pro: UD150B-40 / UD150V-40 / UD150L-40 / UD150L-40E/D150BC-40 / D150VC-40 / D150LC-40 / D150LC-40E/D150BC-41 / D150VC-41 / D150LC-41; Digital Radiography System RADspeed safire: UD150B-40 / D150BC-40; Radiography System RADspeed fit: N/A (Integrated in the system); X-RAY R/F SYSTEM SONIALVISION safire17:UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION safire Plus: UD150B-40 / D150BC-4;0 X-RAY R/F SYSTEM SONIALVISION safire: UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION VERSA: UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM FLUOROspeed 300:UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM SONIALVISION G4:D150BC-40S; X-RAY R/F SYSTEM FLUOROspeed X1:D150BC-40S — Shimadzu Medical Systems

When
· day precision
Where
Long Beach, CA, US
Category
recall · device_recall

Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

Involved actors & entities

People, organizations and places machine-extracted from the source reporting — they power search and the correlation graph. Extracted automatically, so they can include noise, especially on events still marked unverified.

Sources

Defaxon links out to the original reporting and never republishes article text.

Correlated events

Computed by the Defaxon correlation engine — linked by shared actors, co-location, and temporal proximity. Scored hypotheses, never causal claims.

No correlated events found in the current window. As more events arrive, connections form automatically.

← Back to the live map