Device recall: Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays. — The Binding Site Group, Ltd.
- When
- · day precision
- Where
- Birmingham, N/A, GB
- Category
- recall · device_recall
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.
Involved actors & entities
People, organizations and places machine-extracted from the source reporting — they power search and the correlation graph. Extracted automatically, so they can include noise, especially on events still marked unverified.
Sources
Defaxon links out to the original reporting and never republishes article text.
Correlated events
Computed by the Defaxon correlation engine — linked by shared actors, co-location, and temporal proximity. Scored hypotheses, never causal claims.
No correlated events found in the current window. As more events arrive, connections form automatically.