Device recall: Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004 — Fresenius Kabi USA, LLC
- When
- · day precision
- Where
- North Andover, MA, US
- Category
- recall · device_recall
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Involved actors & entities
People, organizations and places machine-extracted from the source reporting — they power search and the correlation graph. Extracted automatically, so they can include noise, especially on events still marked unverified.
Sources
Defaxon links out to the original reporting and never republishes article text.
Correlated events
Computed by the Defaxon correlation engine — linked by shared actors, co-location, and temporal proximity. Scored hypotheses, never causal claims.
No correlated events found in the current window. As more events arrive, connections form automatically.