Device recall: Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039); — PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

When
· day precision
Where
Best, NL
Category
recall · device_recall

It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

Involved actors & entities

People, organizations and places machine-extracted from the source reporting — they power search and the correlation graph. Extracted automatically, so they can include noise, especially on events still marked unverified.

Sources

Defaxon links out to the original reporting and never republishes article text.

Correlated events

Computed by the Defaxon correlation engine — linked by shared actors, co-location, and temporal proximity. Scored hypotheses, never causal claims.

No correlated events found in the current window. As more events arrive, connections form automatically.

← Back to the live map